Case study · Interface ergonomics

When several alarms go off, is the most urgent one still the most visible?

When an operator overlooks an alarm, we call it human error. In human factors, we call it a design error: the person did not fail, the system was designed without regard for human limits.

  • Cognitive ergonomics
  • Visual attention modelling
  • IEC 62366-1
  • IEC 60073

Published 16 November 2026

Full report available on request

Try it. Five alarms are active on this patient monitoring screen. Only one signals an urgent, life-threatening danger. Where do your eyes go?

Screen of a central station monitoring eight patients, with five alarm banners in the same format: one red, two yellow, two cyan. The same screen under a heat map: the hottest areas are the yellow and cyan banners, the red banner barely registers. 5thLife-threatening 1st 2nd 3rd 4th

The life-threatening alarm, in red, comes last. In this example, a technical alert draws the eye more than the life-threatening alarm.

Context

Hospitals, where alarms are everywhere

According to a review of 72 publications, 80 to 99% of the alarms raised by medical devices in hospitals are false or clinically insignificant [1]. Over time, staff get used to them and the risk of ignoring one grows: this is known as alarm fatigue.

The usual response is to reduce the number of alarms. We looked at something a manufacturer controls directly: their visibility. When several alarms are displayed at once, is the most serious one still the most visible to the operator?

The problem

The more alarms there are, the less you see the most serious one

With a single alarm, the hottest area of the screen is the red banner of the life-threatening alarm. With five alarms, the hottest areas are the four other banners; the life-threatening alarm barely registers.
Where the eye lands, from blue (little) to red (a lot), according to a visual attention model. The life-threatening alarm is the same on both screens.

With a single alarm, all is well. The life-threatening alarm is the most visible.

With four more alarms, it loses 41% of its visibility. It has not changed, yet it becomes less visible than the other four.

In practice: when several alarms trigger at once, the interface does not steer attention to the most urgent one. On the contrary, the screen draws the eye first to a secondary, non-urgent alarm, and only then to the life-threatening one. In real conditions, this can lead to a life-threatening emergency being overlooked or treated late.

The cause

Three design choices behind the problem

  1. Red is the darkest colourOn a dark background, yellow and cyan stand out more. The colour that means “most serious” is the one that stands out least.
  2. All alarms share the same formatSame size, same position, same typography. Severity is coded by colour alone.
  3. The screen does not adaptOne alarm or five, the display is the same. Nothing does the sorting for the clinician.

Research on visual perception explains the result: a unique item is spotted at a glance, but as soon as several look alike, they have to be checked one by one [2].

Our method

An audit grounded in standards and a visual attention model

  1. FrameFour conditions, with a growing number of alarms on screen.
  2. Measure44 ergonomic criteria, and the visibility of each alarm quantified by a model.
  3. AnalyseResult cross-checked with two other models.
  4. RecommendRecommendations, interface redesign, then a before/after comparison with the same measure.

The interface was assessed against established ergonomic criteria [3, 4] and checked against two international standards: IEC 62366-1, which governs the usability engineering of medical devices, and IEC 60073, which sets the coding rules for colours and signals [5, 6].

The visibility of each alarm is estimated by our visual attention model. It combines two published methods: the spectral residual, which picks up what stands out by its structure (a sharp edge, an isolated shape) [7], and global colour contrast, which picks up what stands out by its colour [8]. Colour weighs 60% in the model, because it is the first signal of an alarm. The computation is deterministic: the same image always gives the same values. The method is fast, inexpensive and applicable from the mock-up stage, before any user testing.

The solution

Making urgency more visible

Three changes to the interface. The elements and their layout on screen stay the same.

  1. Solid colour for life-threatening danger onlyLess serious alarms become simple outlines. What shines most is what matters most.
  2. One shape and one size per severity levelA pictogram and a format specific to each level, readable even without colour.
  3. A screen that does the sortingAs soon as several alarms are active, a count appears and stable patients recede.

Before on the left · After on the right

Before: five alarm banners in the same format, red, yellow and cyan. After the redesign: only the life-threatening alarm keeps a solid red fill, wider, with a pictogram; the other alarms are outlined and stable patients are dimmed.

Drag the slider. Same patients, same alarms, same situation.

The result

With our redesign, the life-threatening alarm goes from least visible to most visible

Visibility of each alarm, with five alarms on screen

Relative to the screen average. The longer the bar, the more the alarm draws the eye.

Before

Life-threatening×4.8
Medium severity×10.6
Medium severity×10.3
Technical alert×8.8
Technical alert×8.5

After

Life-threatening×10.5
Medium severity×3.7
Medium severity×2.3
Technical alert×1.0
Technical alert×0.7
  • It is 2.2 times more visible than on the original screen, in the same situation.
  • It no longer drops when other alarms appear. The 41% loss is gone.
  • Visibility now follows severity: life-threatening first, then medium severity, then technical alerts.

No information was removed from the screen; only the way alarms are displayed has changed.

For your product

What this study means for you

  • A critical interface should be tested in its most degraded scenarios. With a single alarm, the life-threatening alarm is indeed the most visible. The flaw only appears when several alarms arrive at once, which is when the risk is highest.
  • A colour code is not enough to show what is serious. Brightness, size, shape and context matter just as much.
  • This kind of flaw can be caught early, before any user testing, without mobilising clinicians or equipment.
  • For a medical device, this audit fits into formative evaluation and feeds the usability engineering file.

The principle goes beyond healthcare: control rooms, cockpits, dashboards, applications and software. Wherever a priority signal has to stand out from a busy interface, the method can apply.

Next steps

From prediction to measurement

This study predicts where the eye lands using a computational model of visual attention. It does not replace a study with participants: the next step is to verify these results with an experimental protocol, in the field.

  • Eye trackingMeasuring the clinician’s gaze (fixations, saccades) across the different versions of the interface.
  • PsychophysiologyMental workload, stress and the operator’s physiological response to the interface.
  • UX researchTask analysis, cognitive walkthroughs and interviews with clinicians.

Full report available on request

Deliverables of this type of audit: audit report44-criteria gridbefore/after visibility mapsredesign mock-upsreproducible measurementsprotocol for the next step

We test your interface when everything happens at once.

Tell us about your project

References

  1. Cvach, M. (2012). Monitor alarm fatigue: an integrative review. Biomedical Instrumentation & Technology, 46(4), 268–277. doi:10.2345/0899-8205-46.4.268
  2. Treisman, A. M. & Gelade, G. (1980). A feature-integration theory of attention. Cognitive Psychology, 12(1), 97–136. doi:10.1016/0010-0285(80)90005-5
  3. Bastien, J. M. C. & Scapin, D. L. (1993). Ergonomic criteria for the evaluation of human-computer interfaces. INRIA Technical Report no. 156. inria-00070012
  4. Nielsen, J. (1994). Heuristic evaluation. In J. Nielsen & R. L. Mack (Eds.), Usability Inspection Methods. Wiley.
  5. IEC 62366-1:2015. Medical devices – Part 1: Application of usability engineering to medical devices.
  6. IEC 60073:2002. Basic and safety principles for man-machine interface, marking and identification – Coding principles for indicators and actuators. webstore.iec.ch
  7. Hou, X. & Zhang, L. (2007). Saliency detection: a spectral residual approach. IEEE CVPR. doi:10.1109/CVPR.2007.383267
  8. Cheng, M.-M., Mitra, N. J., Huang, X., Torr, P. H. S. & Hu, S.-M. (2015). Global contrast based salient region detection. IEEE Transactions on Pattern Analysis and Machine Intelligence, 37(3), 569–582. doi:10.1109/TPAMI.2014.2345401
  9. Montabone, S. & Soto, A. (2010). Human detection using a mobile platform and novel features derived from a visual saliency mechanism. Image and Vision Computing, 28(3), 391–402. doi:10.1016/j.imavis.2009.06.006

Study conducted by Nox Science. The device manufacturer is not named; it neither commissioned nor endorsed this analysis. The screens shown are anonymised reconstructions made by Nox Science; the patients and values displayed are fictitious. The redesign is a study mock-up and does not constitute clinical guidance.

Notes on method. The screens are reconstructions faithful to the structure and coding of the product. The result was cross-checked with two published models, which do not take colour into account [7, 9]: all of them rank the life-threatening alarm last on the original screen, and all of them confirm that the loss disappears after the redesign. Only our model places it first; confirming that requires an eye-tracking measurement. The full computation is described in the report (in French), available on request.